Electrolyte Management System
IS9024
Fills and empties large cans during enamel rating
The IS9024 Electrolyte Management System is a semi-automatic system that fills and empties cans, drums and pails with electrolyte during enamel rating. It removes the hand-pouring step that causes most of the spills, and most of the wasted time, on a large-container test.
It is used together with the IS9015 Can Stand. The filling tube holder mounts into the IS9015 arm, and the filling tube, main probe and level probe are refitted into that arm once the holder is secured, so the stand keeps doing the positioning while the IS9024 handles the electrolyte.
Made from specially engineered plastic, the tank and base are built for a factory environment with minimal maintenance. A spillage exit hole in the base routes any overflow to a tray or a hose, and the unit is moved by pushing it on its wheels rather than by lifting the tank.
Electrolyte management system features
| Features |
|---|
| Semi-Automatic Filling and Emptying – the can under test is filled and drained by the system, so the operator never has to lift or pour a heavy container of electrolyte. |
| Three-Speed Filling – three selectable fill speeds cover a wide range of container sizes, from roughly 0.5 litres up to 9 litres and beyond. |
| Remote Control Unit – fill, drain and speed selection are driven from a handheld control unit, with a separate LED pattern confirming each function. |
| Programmable Timing – fill and drain times can be taught per can size and are retained after a power cycle, so a batch of identical cans runs the same way every time. |
Benefits of the IS9024 electrolyte management system
| Benefits |
|---|
| Saves Time – filling and emptying run automatically instead of by hand, which shortens every cycle on a large-can test. |
| Easier With Large Containers – drums and pails that are awkward to fill and empty manually are handled by the system. |
| No Heavy Lifting – nobody has to lift or carry a container filled with electrolyte. |
| Less Risk of Spills – electrolyte is moved through the filling tube rather than poured, and a spillage exit hole in the base routes any overflow to a tray or hose. |
Electrolyte management system spare parts
| Spare Part | Part Number |
|---|---|
| IS9024 Electrolyte Management System | 390-51600-98 |
| IS9024 Tube Holder | 113-51642-91 |
Electrolyte management system specifications
| Specifications | Value |
|---|---|
| Tank Capacity | 18.9 litres |
| Maximum Size of Test Can | 14 litres |
| Minimum Electrolyte in Tank | 5 litres |
| Supply Voltage | 24 V DC ±10% |
| Supply Current | Typically 5 A (an 8 A power supply is recommended and is not included) |
| Weight (Empty Tank) | 12 kg |
| Weight (Full Tank) | Approximately 31 kg |
| Operating Temperature | 0–45 °C (32–113 °F) |
| Humidity | 0–90%, non-condensing |
| Dimensions | 525 mm x 240 mm x 640 mm |
| Used With | Innosen IS9015 Can Stand and IS810 EnamelPro™ Enamel Rating Gauge |
Accessories
Frequently Asked Questions
What does the IS9024 Electrolyte Management System do?
Which can sizes can the IS9024 handle?
Does the IS9024 need the IS9015 Can Stand?
What does semi-automatic filling and emptying mean on the IS9024?
Why do the pumps need priming, and how is it done?
What power supply does the IS9024 need?
How is the IS9024 maintained?
What should I check if the IS9024 will not fill or empty a can?
Complete Enamel Rating Line
Seamless integration with the IS810 EnamelPro™ ecosystem
IS810 EnamelPro™ Enamel Rating Gauge
The semi-automatic enamel rating gauge at the heart of the line. Reverse-polarity defect localization and live logging to Innolytix®.